|
Hem
>
Föredrag och Posters
>
Nuklearmedicin
>
nm3
|
Föregående
Nästa
|
|
|
|
|
|
Hsu BL, Ljungberg M, Case JA, Bateman TM and Cullom SJ
Cardiovascular Imaing Technologies, Kansas City, Missouri and Medical Radiation hysics, Lund
University, Lund
Objectives : Spectral reconstruction previously has been demonstrated to improve the energy
resolution (ER) of gamma camera and then utilized for scatter correction (SC) in cardiac SPECT
(Hsu et al, ASNC 2004). We validated the IESR algorithm on energy data of NCAT phantom
generated from Monte Carlo simulation (MCS) using the SIMIND code.
Methods: IESR assumed the energy spectrum (ES) of gamma camera is a linear combination of
energy components (EC) presented by photon counts contributed from specific energies. It
employed the energy impulse responses (EIR) of system (50-200 keV) to model the response of
camera to the energies of EC for the probability matrix. EC in the spectrum were then iteratively
reconstructed with MLEM method. EIRs with 1 keV sampling interval were provided from MCS of
Cardio60 SPECT system (Philips, Milpitas CA) with ER=10.5% at 140 keV. ES of “in-air” point
sources of Tc99m and Tl- 201 were reconstructed and compared with MCS results. IESR was
then applied to MCS cardiac NCAT phantom projections. Integrated reconstructed spectrum
(IRS) defined by summing ES of 64*64 pixels was compared to the summed ES simulated for
hypothetically ultra fine ER=2%. SC was then performed for each pixel of projection by only
including the EC near photopeaks.
Results : Using 500 iterations (i), IESR effectively separated 70 and 81 keV peaks of Tl- 201
agreed with the ES simulated for ER=2% with minimal differences (Δ) of FWHM (Δ70=1.1 keV,
Δ80=1.75 keV) and FWTM (Δ70=2.2 keV, Δ80=2.7 keV) for the two peaks. For mono energetic
photopeaks of 135 and 167 keV of Tl-201 and 140 keV of Tc99m, the Δ to MCS were overall
less than 2.5 keV. For NCAT cardiac phantom, IESR with i=5000 produced IRS matched with
MCS spectrum for ER=2% with (3.2±1.2)% Δ of counts per keV for Tc99m and (8.3±4.2)% for
Tl-201. Using IESR for SC, the contrast of wall to cavity was improved 25.4% for Tc99m
phantom and 51.2% for Tl-201 with only a slight increase of image noise.
Conclusions : IESR is a physical approach to improve the energy resolution of gamma camera
with the capability of reconstructing energy spectrum matched with 2% energy resolution.
Scatter correction with IESR enhanced the image contrast of myocardial perfusion images of
NCAT phantom generated with Monte Carlo simulation.
|
|
|
|
|
|